Analýza mikroplastů v pivu metodami vibrační spektrometrie
DOI:
https://doi.org/10.54779/ccsss20260335Klíčová slova:
mikroplasty, FTIR mikroskopie, pivo, siMPleAbstrakt
This study evaluates the suitability of Fourier-transform infrared (FTIR) microscopy for detecting microplastics in commercial beers and brewing intermediates. Samples were isolated by vacuum filtration using 0.2 µm aluminum oxide membrane filters and subsequently analysed in transmission mode with an FTIR microscope equipped with a focal plane array (FPA) detector. To improve the signal-to-noise ratio and reduce the large data volume generated during hyperspectral imaging, a 4×4 pixel binning approach was applied. The resulting hyperspectral datasets were processed using the automated siMPle software. The results revealed notable challenges in matrix removal. Treatment with Fenton's reagent caused mechanical instability of the filters and the formation of an infrared-opaque crust, which complicated the subsequent analysis. Despite these limitations in sample preparation, the optimized analytical workflow enabled the detection of microplastic contamination, predominantly polypropylene particles, even in early brewing stages such as sweet wort and hopped wort. Overall, the integration of FTIR mapping, pixel binning, and automated data processing represents a time-efficient and robust strategy for monitoring microplastics in food production systems.